Product overview: Fluke 753 Documenting Process Calibrator
Fluke 753 Documenting Process Calibrator helps you work smarter and faster
The Fluke 753 is a powerful multifunction documenting calibrator that lets you download procedures, lists, and instructions created with software, and upload data for printing, archiving, and analysis. You will find it does the work of several tools, with simultaneous source and measurement capabilities for all common process parameters. It sources, simulates, and measures pressure, temperature, and electrical signals in one rugged, hand-held calibration tool. It also automates calibration procedures, captures data for documentation, and helps you meet rigorous standards like ISO 9000, FDA, EPA, and OSHA regulations.
Key features
Technical Data |
|
|
Data log functions |
Measure functions |
Voltage, current, resistance,
frequency, temperature, pressure |
|
Reading rate |
1, 2, 5, 10, 20, 30, or 60
readings/minute |
|
Maximum record length |
8000 readings (7980 for 30 or 60
readings/minute) |
Ramp functions |
Source functions |
Voltage, current, resistance,
frequency, temperature |
|
Rate |
4 steps/second |
|
Trip detect |
Continuity or voltage (continuity
detection not available when sourcing current) |
Loop power function |
Voltage |
Selectable, 26 V |
|
Accuracy |
10%, 18 V minimum at 22 mA |
|
Maximum current |
25 mA, short circuit protected |
|
Maximum input voltage |
50 V DC |
Loop power function |
Voltage |
Selectable, 26 V |
|
Accuracy |
10%, 18 V minimum at 22 mA |
|
Maximum current |
25 mA, short circuit protected |
|
Maximum input voltage |
50 V DC |
Mechanical and General Specifications |
|
|
Size |
136 x 245 x 63 mm (5.4 x 9.6 x 2.5
in) |
|
Weight |
1.2 kg (2.7 lb) |
|
Batteries |
Internal Battery Pack Li-ion 7.2 V,
4400 mAh, 30 Wh |
|
Battery life |
> 8 hours typical |
|
Battery replacement |
Via snap-shut door without opening
calibrator no tools required |
|
Side port connections |
Pressure module connector |
|
|
USB Connector to interface to your PC |
|
|
Connection for optional battery
charger/eliminator |
|
Data storage capacity |
1 week of calibration procedures
results |
|
90 day specifications |
The standard specification interval
for the 750 Series are 1 and 2 years. |
|
|
Typical 90 day measurement and source
accuracy can be estimated by dividing the one year % of reading or % of
output specifications by 2. |
|
|
Floor specifications, expressed as %
of full scale or counts or ohms remain constant. |
Models: Fluke 753 Documenting Process Calibrator
Fluke 753Fluke 753 Documenting Process Calibrator
Includes: